Example: Influence of datum of pressure measurement in the triaxial test
In the triaxial test, pressure transducers can be used to measure cell pressure, back-pressure, pore pressure in the base pedestal and pore pressure in the mid-plane of the test specimen. In the hydraulic stress path triaxial cell (such as the Bishop and Wesley, 1975 design), lower chamber pressure will also be measured. In this extreme example, there will be up to five pressure transducers. Naturally, there will be small differences between measurements of the same pressure by these transducers. This is because they will have slightly different accuracies (their specified deviation from a standard value). This is quite normal and should be taken into account when interpreting test results because we do not have any control over these inherent discrepancies. We do, however, have control over setting the common zero or datum of pressure measurement. This is essential for all four or five pressure measuring systems (i.e. the transducers and their associated analogue-digital conversion) to ensure that we measure pressure from the same “base line” of total head.
Fig. 11 Set-up schematic diagram for setting the datum of pressure measurement to the mid-height (or base if preferred) of the test specimen in the triaxial test when various pressure sensors are at different elevations
The first thing to do is to set up the datum of pressure measurement. Normally this will be an elevation equal to the mid-height of the test specimen. Probably the best way of doing this is to connect a short length (say 300mm) of small bore nylon tubing with an open end to the back pressure connector of the cell. This is the connector to the top drain. Fill the cell with water. There will not be a test specimen in place for this procedure. Any mid-plane pressure transducer will be laid in the bottom of the inside of the cell together with the top cap and back pressure connection to it. |
Consider the triaxial set-up shown in Fig. 11. The cell pressure and back pressure are applied using GDS digital pressure controllers. These might be mounted on a bench alongside the triaxial frame or mounted on the wall near the frame. Either way they will probably be at different elevations to the triaxial cell. Fill the cell with water. Apply a small positive cell pressure using the cell pressure controller. Open the valve to the back pressure line. Water will flow out of the cell into the short tube. Stop the controller from pumping when the tube is full of water and drips out of the open end. Fix the open end of the water-filled tube at an elevation corresponding to the mid-height of the test specimen (or the base of the test specimen if you prefer to use that elevation as the datum).
Now the water in the cell is at a pressure corresponding to this elevation head. Connect the back-pressure controller to the base pedestal pore water port and open the valve. Now the cell pressure controller, back-pressure controller and pore pressure transducers all share the same pressure head set by the external tube and would deliver the same zero pressure as at the mid-sample elevation. The displays of the pressure controllers and the pore pressure transducers can be zeroed at this time. Now all four (or five) displays of pressure are zeroed to the same datum of pressure measurement. For the stress path cell the lower chamber pressure controller can be zeroed at this time also. The zeroing process is a soft (as in software) zero offset – the hardware zero is not touched.
What if the different pressure transducers were zeroed at different times during the set up of the test with corresponding discrepancies? For example, consider the set up shown in Fig.11. Remember that 10kPa corresponds to a head of very nearly 1m of fresh water. If the zero offset for the pore pressure transducer is applied, the actual pressure at the pore pressure transducer will be 3kPa (this is the head of water) but because a zero offset is applied (by zeroing the corresponding channel of the logging system) this will be read as 0kPa. But consider now if the zeros had been set on the cell pressure and back pressure controllers when they were open to the atmosphere before being connected to the cell pressure and back pressure lines. Then the pressure datum for each of them will be the respective elevations of the outlets of the controller pressure cylinders. Now, for the set up shown in Fig.11, the cell pressure controller will measure (3 + 3 = 6kPa), the back pressure controller will measure (3 + 3 + 3 = 9kPa), but the pore pressure transducer will read zero because it has been zeroed in this condition. Therefore the two controllers will have a 6kPa or 9kPa difference from the pore pressure measurement due to their different zeros of pressure measurement. To overcome this discrepancy, it is essential to zero all devices at the same time using the same datum as described in this section.
|